Experimental use of an illuminated and magnetic tip, flow-directed pulmonary artery catheter for lung resection
Autore:
Fulton, J; Conlan, AA; Soller, BB; Favreau, J; Hsi, C; Kim, H;
Indirizzi:
Univ Massachusetts, Sch Med, Dept Surg, Mem Med Ctr, Worcester, MA 01605 USA Univ Massachusetts Worcester MA USA 01605 ed Ctr, Worcester, MA 01605 USA Univ Massachusetts, Sch Med, Div Cardiothorac Surg, Worcester, MA 01605 USA Univ Massachusetts Worcester MA USA 01605 c Surg, Worcester, MA 01605 USA
Titolo Testata:
SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES
fascicolo: 2, volume: 11, anno: 2001,
pagine: 107 - 111
SICI:
1051-7200(200104)11:2<107:EUOAIA>2.0.ZU;2-I
Fonte:
ISI
Lingua:
ENG
Soggetto:
ASSISTED THORACIC-SURGERY; LOBECTOMY; EXPERIENCE; COMPLICATIONS; THORACOSCOPY;
Keywords:
balloon catheter; endovascular control device; lobectomy; magnetic tip; pulmonary artery; thoracoscopic;
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Clinical Medicine
Citazioni:
11
Recensione:
Indirizzi per estratti:
Indirizzo: Conlan, AA Univ Massachusetts, Sch Med, Dept Surg, Mem Med Ctr, Mem Campus,67 BelmontSt, Worcester, MA 01605 USA Univ Massachusetts Mem Campus,67 Belmont St Worcester MA USA 01605
Citazione:
J. Fulton et al., "Experimental use of an illuminated and magnetic tip, flow-directed pulmonary artery catheter for lung resection", SURG LA E P, 11(2), 2001, pp. 107-111

Abstract

The authors performed a study to design and evaluate a device giving the surgeon added protection against pulmonary artery injury in a closed thoracic compartment. Eleven swine were used for a total of 12 lobectomies. A light-bearing, magnetic-tipped, modified Swan-Ganz catheter was passed through the jugular vein into the left pulmonary artery. By using magnetic guidancethrough a port site and balloon inflation, cessation of flow in the pulmonary artery was documented by Doppler. Twelve video-assisted lobectomies were performed. Nine of 12 (75%) lobectomies were completed successfully by using the magnetic-tipped, illuminated balloon catheter. In 4 cases, the catheter displaced because of the short left main pulmonary artery in swine. Inone case, the balloon was recovered, and successful lobectomy was performed. The use of a magnetic-tipped, flow-directed device may provide an effective means of endovascular control during thoracoscopic pulmonary arterial dissection.