Transcatheter retrieval and repositioning of an Amplatzer device embolizedinto the left atrium
Autore:
Peuster, M; Boekenkamp, R; Kaulitz, R; Fink, C; Hausdorf, G;
Indirizzi:
Hannover Med Sch, Dept Pediat Cardiol & Pediat Intens Care, D-3000 Hannover, Germany Hannover Med Sch Hannover Germany D-3000 Care, D-3000 Hannover, Germany
Titolo Testata:
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS
fascicolo: 3, volume: 51, anno: 2000,
pagine: 297 - 300
SICI:
1522-1946(200011)51:3<297:TRAROA>2.0.ZU;2-S
Fonte:
ISI
Lingua:
ENG
Soggetto:
DOUBLE-DISC DEVICE; SEPTAL OCCLUDER; OCCLUSION DEVICES; CLOSURE; DEFECTS; EXPERIENCE;
Keywords:
atrial septal defect; closure devices; Amplatzer; embolization;
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Clinical Medicine
Citazioni:
10
Recensione:
Indirizzi per estratti:
Indirizzo: Peuster, M Hannover Med Sch, Childrens Hosp, Dept Pediat Cardiol & Pediat Intens Care, D-30623 Hannover, Germany Hannover Med Sch Hannover Germany D-30623 3 Hannover, Germany
Citazione:
M. Peuster et al., "Transcatheter retrieval and repositioning of an Amplatzer device embolizedinto the left atrium", CATHET C IN, 51(3), 2000, pp. 297-300

Abstract

A transcatheter technique is described for stabilization and retrieval of an embolized Amplatzer device, which was inadvertently deployed in the leftatrium of a 5-year-old girl with hemodynamically significant atrial septaldefect within the fossa ovalis. Since surgery was not available for immediate device retrieval, transcatheter means were chosen to prevent embolization of the device into the mitral valve. After stabilizing the device by creating a guidewire circuit through the wire mesh of the right atrial disk, the Amplatzer device was repositioned through the interatrial defect by snaring the microscrew. No residual shunting and perfect device position was recorded during Doppler echocardiography directly after the procedure as wellas 1, 2, and 120 days after the intervention. Cathet. Cardiovasc. Intervent. 51:297-300, 2000. (C) 2000 Wiley-Liss, Inc.