Graft-wall endoleak 18 months after successful endoluminal AAA repair
Autore:
Midorikawa, H; Hoshino, S; Iwaya, F; Igari, T;
Indirizzi:
Fukushima Med Univ, Sch Med, Dept Cardiovasc Surg, Fukushima, Japan Fukushima Med Univ Fukushima Japan pt Cardiovasc Surg, Fukushima, Japan
Titolo Testata:
JOURNAL OF ENDOVASCULAR SURGERY
fascicolo: 3, volume: 6, anno: 1999,
pagine: 251 - 255
SICI:
1074-6218(199908)6:3<251:GE1MAS>2.0.ZU;2-X
Fonte:
ISI
Lingua:
ENG
Soggetto:
3-YEAR EXPERIENCE; STENT-GRAFT; ENDOGRAFTS;
Keywords:
endovascular grafts; endoleak; stent-graft rupture; complications;
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Clinical Medicine
Citazioni:
8
Recensione:
Indirizzi per estratti:
Indirizzo: Midorikawa, H 1 Hikarigaoka, Fukushima 9601295, Japan 1 Hikarigaoka Fukushima Japan 9601295 shima 9601295, Japan
Citazione:
H. Midorikawa et al., "Graft-wall endoleak 18 months after successful endoluminal AAA repair", J ENDOVAS S, 6(3), 1999, pp. 251-255

Abstract

Purpose: To describe a case of graft-walt endoleak 18 months after successful endoluminal repair of an abdominal aortic aneurysm (AAA). Methods and Results: A 71-year-old man with infrarenal AAA was successfully treated with an endoluminal aortomonoiliac graft and femorofemoral crossover bypass with surgical ligation of the right external iliac artery. The stent-g raft was made from 2 Gianturco Z-stents and a tapered thin-walled (0.1-mm) Dacron graft. Eighteen months after endografting, the patient complained of a pulsatile abdominal mass. Angiography and computed tomography showed graft-wall endoleak. Aneurysmectomy was performed, and the aneurysm wassuccessfully replaced with a Y-shaped knitted Dacron graft. A hole in the graft wall was found 3 cm from the proximal edge of the stent-graft. Conclusions: This case suggests that the use of thin-walled graft materialin endografts may not be sufficiently durable.