Catalogo Articoli (Spogli Riviste)

OPAC HELP

Titolo:
THE USE OF A PORTABLE REFRACTOMETER TO MEASURE DESFLURANE
Autore:
CHEAM EWS; LOCKWOOD GG;
Indirizzi:
HAMMERSMITH HOSP,ROYAL POSTGRAD MED SCH,DEPT ANAESTHESIA,DU CANE RD LONDON W12 0HS ENGLAND
Titolo Testata:
Anaesthesia
fascicolo: 7, volume: 50, anno: 1995,
pagine: 607 - 610
SICI:
0003-2409(1995)50:7<607:TUOAPR>2.0.ZU;2-4
Fonte:
ISI
Lingua:
ENG
Keywords:
MEASUREMENT TECHNIQUES, REFRACTOMETER, INFRARED ABSORPTION; ANESTHETICS, VOLATILE, DESFLURANE;
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Science Citation Index Expanded
Citazioni:
4
Recensione:
Indirizzi per estratti:
Citazione:
E.W.S. Cheam e G.G. Lockwood, "THE USE OF A PORTABLE REFRACTOMETER TO MEASURE DESFLURANE", Anaesthesia, 50(7), 1995, pp. 607-610

Abstract

The measurement of volatile anaesthetic agents by infrared absorptionis liable to interference by other gases which absorb infrared light. In particular, nitrous oxide may, in principle, interfere with the measurement of desflurane. On the other hand, the refractometer should be immune to such effects. We demonstrated that interactions may occur between desflurane and nitrous oxide with some gas analysers and that the refractometer remains a useful instrument.