REPORT ON THE COOPERATIVE DETERMINATION O F MOLECULAR-WEIGHT AVERAGESOF POLYMERS BY SIZE-EXCLUSION CHROMATOGRAPHY .5. DEPENDENCE OF MOLECULAR-WEIGHT AVERAGES ON DETECTORS AND MOBILE PHASES
Autore:
MORI S; TAKAYAMA S; GOTO Y; NAGATA M; KINUGAWA A; HOUSAKI T; YABE M; TAKADA K; SUGIMOTO T; SHIMIZU M; NAGASHIMA I; HASEGAWA A; SENBA T; OOSHIMA N; MAEKAWA T; SUGITANI H; OOZEKI H; NAKAHASHI K; HIBI K; OOTANI H; NAKAMURA S; SUGIURA K; UMEDA Y; KATSUNO Y; TAGUCHI M;
Indirizzi:
JAPAN SOC ANALYT CHEM,RES COMM POLYMER ANAL,SECT SPECIAL STUDY SEC,SHINAGAWA KU TOKYO 141 JAPAN
Titolo Testata:
Bunseki Kagaku
fascicolo: 1, volume: 46, anno: 1997,
pagine: 81 - 86
SICI:
0525-1931(1997)46:1<81:ROTCDO>2.0.ZU;2-1
Fonte:
ISI
Lingua:
JPN
Keywords:
SIZE EXCLUSION CHROMATOGRAPHY; ROUND ROBIN TEST; POLYSTYRENE; POLY(METHYL METHACRYLATE); DEPENDENCE OF MOLECULAR-WEIGHT AVERAGES ON DETECTORS AND MOBILE PHASES;
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Science Citation Index Expanded
Citazioni:
5
Recensione:
Indirizzi per estratti:
Citazione:
S. Mori et al., "REPORT ON THE COOPERATIVE DETERMINATION O F MOLECULAR-WEIGHT AVERAGESOF POLYMERS BY SIZE-EXCLUSION CHROMATOGRAPHY .5. DEPENDENCE OF MOLECULAR-WEIGHT AVERAGES ON DETECTORS AND MOBILE PHASES", Bunseki Kagaku, 46(1), 1997, pp. 81-86

Abstract

The effects of the differences in detectors and mobile phases on the calculated molecular-weight averages are discussed based on the data obtained from 1st and 2nd round-robin tests. Although the weight-average molecular weights (M(w)) of polystyrene (PS) samples remained unchanged with the detectors, the Dumber-average molecular weights (M(n)) obtained with an ultraviolet absorption detector (UV) were 74 similar to88% of those with a differential refractometer (RI). The fractions ata lower molecular weight range of the differential molecular-weight distributions, calculated using UV chromatograms, were higher than those based on RI chromatograms, which caused the M(n) obtained by a UV detector to have smaller values than those by an RI detector. Peak broadening due to the cell structure or the difference in the molecular-weight dependence of the response coefficients of UV and RI detectors were negative as the main reason. The most probable assumption was that the end-structure of the PS samples affected the increase in the UV absorption. When a comparison of the M(n) values between laboratories is required, the type of detector for size-exclusion chromatography must be the same. The values of M(w) and M(w) for poly(methyl methacrylate), obtained with chloroform mobile phase, were 5 similar to 25% higher than those with the tetrahydrofuran mobile phase. PS standards were used for the calibration; in this case, the change in the values of M(n)and M(w) for other types of polymers than PS using different mobile phases had to be taken into account.