400MHZ 89MM ACTIVELY SHIELDED HIGH-RESOLUTION NMR MAGNET/
Autore:
TEODORESCU R; COLVIN J; SHOTZMAN G; VICK T; ZEIGLER J; LIANG G; CROWE L; ABSHIRE C; MANN T; PARAKH Z; SHEN W;
Indirizzi:
HOUSTON ADV RES CTR,ASTROPARTICLE PHYS GRP THE WOODLANDS TX 77381
Titolo Testata:
IEEE transactions on magnetics
fascicolo: 4, volume: 32, anno: 1996,
parte:, 1
pagine: 2647 - 2650
SICI:
0018-9464(1996)32:4<2647:48ASHN>2.0.ZU;2-4
Fonte:
ISI
Lingua:
ENG
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Science Citation Index Expanded
Science Citation Index Expanded
Citazioni:
3
Recensione:
Indirizzi per estratti:
Citazione:
R. Teodorescu et al., "400MHZ 89MM ACTIVELY SHIELDED HIGH-RESOLUTION NMR MAGNET/", IEEE transactions on magnetics, 32(4), 1996, pp. 2647-2650

Abstract

The first actively shielded high resolution NMR magnet was developed and successfully demonstrated, The magnet has excellent field homogeneity in a very large volume and lineshape characteristics exceed the specifications of the standard magnets, Shielding performance is also outstanding, with the 5 G line located only 1 m radially and 1.5 m axially from the magnet center. We are reporting the magnet specifications,details of the design and fabrication, magnet training, field mappingand lineshape results.