Catalogo Articoli (Spogli Riviste)

OPAC HELP

Titolo:
A NEW DOUBLE-LUMEN BALLOON CATHETER FOR RETROGRADE CEREBRAL PERFUSIONVIA JUGULAR-VEIN CANNULATION
Autore:
DOI K; SATOH S; HIRAI J; NIU S; KANDA K; OKANO T; MASUDA S; OKA T;
Indirizzi:
KYOTO PREFECTURAL UNIV MED,DEPT SURG 2,KAMIKYO KU KYOTO 602 JAPAN KYOTO PREFECTURAL UNIV MED,DEPT SURG 2,KAMIKYO KU KYOTO 602 JAPAN
Titolo Testata:
ASAIO journal
fascicolo: 5, volume: 43, anno: 1997,
pagine: 731 - 735
SICI:
1058-2916(1997)43:5<731:ANDBCF>2.0.ZU;2-U
Fonte:
ISI
Lingua:
ENG
Soggetto:
AORTIC-ARCH; PRESSURE; SURGERY;
Tipo documento:
Article
Natura:
Periodico
Settore Disciplinare:
Science Citation Index Expanded
Citazioni:
14
Recensione:
Indirizzi per estratti:
Citazione:
K. Doi et al., "A NEW DOUBLE-LUMEN BALLOON CATHETER FOR RETROGRADE CEREBRAL PERFUSIONVIA JUGULAR-VEIN CANNULATION", ASAIO journal, 43(5), 1997, pp. 731-735

Abstract

A new catheter for retrograde cerebral perfusion (RCP) was developed that can be used to deliver blood directly into the internal jugular vein (IJV) beyond the venous valves at the jugular-subclavian junction and prevent blood from draining into the lower half of the body. This catheter can be inserted into the IJV via a standard puncture technique by use of a 14 Fr sheath. The catheter shaft has two channels for balloon inflation and blood perfusion, respectively. A balloon for occlusion of drainage veins (superior vena cava and azygos vein) is installed at the catheter tip. Side holes, through which oxygenated cold blood is delivered into the IJV, are located 95 mm from the catheter tip. In a mock circulatory study, the pressure at the perfusion line (16-118 mmHg) increased with the increasing flow rate (0-400 ml/min). In clinical application, under circulatory arrest with profound hypothermia,inflation of the balloon effectively reduced blood drainage into the lower half of the body and, consequently, RCP was successfully performed (flow rate, 300-350 ml/min; pressure at the IJV, 15 mmHg). Because all of these procedures were controlled from outside the operative field, RCP by use of this catheter could be useful in distal arch replacement via left lateral thoracotomy.